|
As shown
in Fig 1(a), two standard SLEDs centered at 750nm and 840nm
are combined together by a 50/50 fiber coupler to form a
broadband source. The measured source spectrum (red) and
interference spectrum (blue) is shown in Fig 1(b).
The source
spectrum is not Gaussian. Hence, there would be sidelobes in
the A-scan signal, which will cause images to blur rather
than increasing the image resolution. Fig 1(c) shows the
result by directly inverse Fourier transforming the
interference signal in Fig 1(b). It is clear that the
deconvolution of the spectrum needs to be performed in order
to achieve a higher resolution.

Fig.1.(a) FD-CP-OCT setup

(b) measured source and interference spectrum

(c) OCT A-scan without deconvolution
Various deconvolution algorithms were applied to obtain
higher resolution in OCT. Here, we used a filter which is
designed to minimize the error between the restored and the
original signal in the mean squared sense, to enhance the
resolution while suppressing the noise. The method is also
referred as Wiener deconvolution .
In OCT, the effectiveness of deconvolution algorithm is
based on the phase continuity between the interference
spectra of the two sources. If the SNR is too low at the
valley between sources, this continuity won’t exist, and the
two sources are absolutely independent and any deconvolution
wouldn’t work well.We quantitatively studied the
effectiveness of deconvolution algorithm as a function of ΔΛ
and SNR. We assume two sources are centered at 800nm ΔΛ/2,
and each with the bandwidth (Full Width Half Maximum) of
35nm. In our simulation, the sample is only a reflectance
layer with a delay of l’ from the reference surface. In this
case, an A-scan OCT signal can be treated as random
variable, with mean l’, and its standard derivation is
proportional to the axial resolution. α, the ratio of
resolution after deconvolution to resolution b, is used as a
measure of the effectiveness the deconvolution algorithm. β
is defined as ΔΛ/δλ , hence α is a function of β and SNR.
Fig 2 shows how α changes under different β and SNR. A
certain combination of β and SNR determines a specific α
value and corresponds to a pixel in Fig2.

Fig.2. α versus SNR and β

Fig 4(a) Fig 4 (b)
Images of cellophane films in Fig 4(a) and (b) are obtained
by using single source and dual source respectively for
comparison. Fig 4(b) shows an improved OCT image which
exhibits a much better axial resolution than the one with
the single source.
Hemoglobin oxygen measurement
The basic experimental setup is
described as above. Interference spectra were recorded for
OCT image reconstruction and SO2 analysis. This can be
expressed as I(k)=η|E0(k)|2∫rs(k,l)cos(2kl)dl, where η is a
constant; rs(k, l) is the depth dependant reflectance; E0(k)
is the electrical field of source output. For homogenous
blood sample, rs(k,l) decays exponentially as a function of
the scanning depth, l. The attenuation coefficient of blood
sample contains scattering coefficient αs and absorption
coefficient αa, which is a function of wavelength as well as
a function of SO2.
rs(k, l)=
rs0exp{-[αs+αa(k)]l} αa(k)=SO2αHbO2(k)+[1- SO2]
αHb(k)
For a biological sample such as chicken embryo, αa, varies
as tissue composition changing at difference depth,
therefore rs(k,l) has a much more complex dependence of l.
2D OCT image and SO2 map are simultaneously obtained by
processing spectra measured at different lateral position.
Fig 5 is the signal processing procedure for a single
A-Scan, similar as in spectroscopic OCT study

Fig. 5. Flow chart of signal processing
R(l’) is the difference of spectrum
between short and long wavelength range. In Fig 6(b), in the
lower image R map and OCT image overlaped. And the upper
curve is obtained by averaging R(l’) along the depth
direction.
In our work, R(l’) does not directly provide this ratio,
nevertheless the technique is effective in identifying the
blood vessels and gives relative value of the SO2 level in
tissue.

Fig 6(a)

Fig 6(b) |